OPTIONS Isolated 9-32 Vdc power Isolated 26-56 Vdc power NEMA-4 splash-proof lens cover Model 2002A-S is a low-cost 4 1/2 digit (±19,999 count) panel meter for indication-only applications that require bridge excitation and resolution down to 1 µV. Zero, span and bridge-balance potentiometers are accessible behind the lens, so that the 2002A-S can be adjusted to read out directly in engineering units for a wide range of bridge inputs or other low-level signals. A ratiometric reference from the bridge excitation eliminates AC line and load regulation errors and reduces overall meter span tempco to 0.01% of reading/°C. Overall meter accuracy is 99.98% at 25°C. The 2002A-S is housed in a short 4.1 in (104 mm) 1/8 DIN case to fit where space is limited. Power options include isolated 9-32 Vdc and 26-56 Vdc. A splash-proof lens cover which meets NEMA-4 requirements allows the 2002A to be used in harsh industrial environments. APPLICATIONS The 2002A-S is ideal as a digital indicator for pressure cells and metal-foil load cells. It can also be used as a low-level, high-impedance process meter with zero and span adjust- ments or as a DC voltmeter with resolution down to 1 µV, as required by 50 mV current shunts. BRIDGE EXCITATION SUPPLY The excitation supply is transformer-isolated to 500 V ac from the rest of the meter and provides up to 30 mA at 10 Vdc of output current. A multiturn potentiometer accessible behind the lens allows adjustment of output voltage from 10 to 24 Vdc. MICROVOLT PREAMPLIFIER The preamplifier provides an input resistance in excess of 1 G without bridge balance. Gains of 1, 10 and 100 are selectable by plug-in jumpers and provide meter resolutions of 100, 10 and 1 µV/count, respectively. Typical offset drift is only 0.3 µV/°C. SPECIFICATIONS ANALOG INPUT
Input resistance: 1 Gohm without bridge balance Bias current: 1 nA typ, 5 nA max Gain attenuator: Adjustable from 0.001 to 1 Bridge-balance range: ±1.5 mV with 350 bridge Zero-adjustment range: 19,999 to +19,999 counts Span-adjustment range: 0 to 19,999 counts REFERENCE Configuration: Differential, ratiometric Internal from excitation supply: 1.0 V with 9.5 kohm source resistance at 10 V excitation External, optional: 1.0 V 50%/+100% with 680 kohm input resistance ACCURACY AT 25°C Maximum Error: ±0.01% of span ±2 counts
Full-scale step response: 1 s Warmup to rated accuracy: 10 min NOISE REJECTION NMR, SIG HI to SIG LO: 80 dB at 50/60 Hz for 20 mV range, 66 dB at 50/60 Hz for 200 mV and 2 V ranges CMR, SIG GND to PWR GND: 120 dB, DC to 60 Hz CMV, SIG GND to PWR GND: 1500 Vp per HV test, 354 Vp per IEC spacing ANALOG-TO-DIGITAL CONVERSION Technique: Dual-slope, average-value Signal integration period: 100 ms Read rate: 2.5/s BRIDGE EXCITATION SUPPLY Output voltage: Adjustable from 10 to 24 Vdc Output current: 30 mA max at 10 Vdc decreasing to 12 mA max at 24 Vdc Line regulation*: 0.01% typ, ±0.04% max for 10% change of AC power voltage Load regulation*: ±0.15% typ, ±0.5% max from zero to max load Ripple at 50/60 Hz: ±0.01% Tempco: ±0.02%/°C *The internal reference is derived from the excitation voltage for ratiometric operation. This eliminates load and line regulation errors and reduces other errors as indicated under accuracy at 25°C. DISPLAY Type: 7-segment, red LED Height: 0.56 in (14.2 mm) Symbols: 1.8.8.8.8 Decimal points: Four positions programmable by jumpers behind lens or at connector, sink 10 mA to SIG GND Overrange indication: Four least-significant digits flash DIGITAL SIGNALS Input signals: Display hold, blanking, display test, decimal point select Level, Input signals: TTL or 5 V CMOS compatible POWER AC voltages: 115 or 230 Vac ±15% AC frequency: 49-440 Hz DC voltages: 9-32 Vdc, isolated to 300 Vp; 26-56 Vdc, isolated to 300 Vp Power consumption: 4.0 W Output Voltages: +4.7 Vdc and 4.7 Vdc ±5%, 10 mA max ENVIRONMENTAL Operating temperature: 0 to 60°C Storage temperature: 40 to +85°C Relative humidity: 95% at 40°C (non-condensing) MECHANICAL Dimensions: Newport DIN1A (1/8 DIN) case (see Mechanical section for drawings) Weight: 15 oz (425 g) Case material: 94V-0 UL-rated polycarbonate Connector for signal, power and excitation: Screw-terminal barrier strip (standard) Connector for display control lines: 36-pin PCB edge connector (optional) HOME / PRODUCT GUIDE / TECHNICAL REFERENCE NEWPORT Electronics, Inc. (714) 540-4914 or Contact Us Copyright 1998, Newport Electronics, Inc. All rights reserved. |